11.1.3 Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
11.1.3.1 Power Connection A.C. Input
the device is designed to operate in the electromagnetic environment specified below.
The customer or the user of the device should ensure that it is used in such environment.
Basic EMC
Phenomenon
standard or test
Electrical fast
transients /
IEC 61000-4-4
bursts
a) l) o)
Surges
a) b) j) o)
IEC 61000-4-5
Line-to-line
Surges
a) b) j) k) o)
Line-to-
IEC 61000-4-5
ground
Conducted
disturbances
IEC 61000-4-6
induced by RF
fields
c) d) o)
Voltage dips
IEC 61000-4-11
f) p) r)
Voltage
interruptions
IEC 61000-4-11
f) i) o) r)
a) The test may be performed at any one power input voltage within the device RATED voltage range. If the device is
tested at one power input voltage, it is not necessary to re-test at additional voltages.
b) All the device cables are attached during the test.
c) Calibration for current injection clamps shall be performed in a 150 Ω system.
d) If the frequency stepping skips over an ISM or amateur band, as applicable, an additional test frequency shall be used in
the ISM or amateur radio band. this applies to each ISM and amateur radio band within the specified frequency range.
28
Immunity test levels
method
±2 kV on contact
100 KHz repetition
frequency
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2kV
3 V
m)
0.15 MHz - 80 MHz
6 V
in the ISM bands
m)
between 0.15 MHz and
80 MHz
80 % AM at 1 KHz
0% Ut; 0,5 cycle g)
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % Ut; 1 cycle e
70 % Ut; 25/30 cycle
Single phase: at 0°
0% Ut; 250/300 cycle
Environment Guidance
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Portable and mobile RF
communication devices
should not be used near any
part of the product, including
cables, except when they
respect the recommended
n)
distances, calculated from the
e)
equation applicable at the
frequency of the transmitter.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
h)
The quality of the network
voltage should be that of a
h)
typical commercial or hospital
environment.
Electromagnetic