Télécharger Imprimer la page

Mectron starlight uno Mode D'emploi Et D'entretien page 32

Masquer les pouces Voir aussi pour starlight uno:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

starlight uno
10
TROUBLESHOOTING
If the equipment appears not to be working correctly, read the instructions again and then check
the table 6 at page 24.
Problem
The charging unit does not
switch on (none of the LEDs
light up).
There is no light emission
when the on/off button of
the starlight uno is pressed
and an acoustic signal is
produced (2 beeps).
An acoustic signal is
produced at the end of the
exposure cycle (2 beeps).
An acoustic signal (3
beeps) is produced during
the exposure cycle and
at the end of the cycle
starlight uno will not allow
any further treatment to be
carried out.
The polymerisation is
insufficient.
An acoustic signal is
produced (4 beeps) and
there is no light emission.
Table 6 – Troubleshooting
24
Possible Cause
The power cable is not
correctly connected.
The power cable is faulty.
The charging unit is out of
order.
Battery discharged.
Battery low.
The thermal protection has
been activated.
The terminal surface of the
tip is soiled.
Hardware failure.
Solution
Connect the cable both to
the charging unit and to the
wall socket.
Replace the power cable.
Contact an authorised
MECtRON technical
assistance centre.
Recharge the battery. See
Chapter 5.3 on page 13.
Recharge the battery. See
Chapter 5.2 on page 13.
It will be possible to use the
equipment only after it has
cooled down. See
Chapter
6.5 on page 17.
See
Chapter 6.4 on page
17.
Contact an authorised
MECtRON technical
assistance centre.

Publicité

loading