7
CLEANING, DISINFECTION AND
STERILISATION
The Table 4 at page 18 is purely indicative.
For the complete cleaning, disinfecting and sterilization procedures of the individual parts, refer
to the paragraphs indicated in the Table 4 at page 18.
CAUTION: All the phases indicated in the following table must be performed, avoiding
using methods not referred to in that table.
! WARNING: The protective sheaths are single use. Each protective sheath must be used for
one application only on a single patient.
Phase
Chapter
Chapter 7.1
on page
I
18
Chapter 7.2.1
II
on page
20
Chapter 7.2.2
III
on page
21
Chapter 7.2.3
IV
on page
21
Chapter 7.2.4
V
on page
21
Table 4 – Cleaning, Disinfection and Sterilisation
7.1
Cleaning and Disinfection of the Handpiece and the
Charging Unit
! WARNING: Switch off the charging unit.
Before carrying out any cleaning and disinfection, disconnect the charging unit from the mains
power supply..
7.1.1
Preparation
• Remove the handpiece from the charging unit;
• Remove the optical protection from the handpiece;
• Remove the protective sheath from the handpiece.
18
Procedure
Manual cleaning with
detergent solution and
disinfecting agent
Immersion into enzymatic
detergent
Check cleaning
Drying
Sterilisation
Charging
Handpiece
unit
X
X
Optical
protection
X
X
X
X

Publicité