CLEANING, DISINFECtION AND StERILISAtION
7.1.2
Necessary Material
• Clean, soft cloths with low fiber-release;
• Detergent solution (pH 6-9);
• Demineralised water;
• Disinfectant agent (Glutaraldehyde, Ghlorhexidine gluconate or Isopropyl alcohol 70%).
7.1.3
Cleaning Method
1.
Clean the surface of the charging unit and the handpiece with a clean, soft and low fibre
release cloth, dampened with a cleaning solution (pH 6-9) prepared according to the
manufacturer's instructions;
2.
Clean the surface of the charging unit and the handpiece with a clean, soft and low fibre
release cloth, dampened with demineralised water to remove all residues of the cleaning
solution;
3.
Dry the surface of the charging unit and the handpiece with a clean, soft and low fibre
release cloth;
4.
If you intend to disinfect, spray the disinfecting agent onto a clean, soft and low fibre
release cloth and clean the surface of the charging unit and of the handpiece.
CAUTION: Do not use as disinfecting agents:
• Very alkaline products (pH > 9);
• Products containing sodium hypochlorite;
• Products containing hydrogen peroxide;
• Products containing abrasive substances;
• Acetone;
• Methylethylketone.
as they can discolour and/or damage the plastic materials.
CAUTION: Contact of liquid with the lamp terminals can cause damage and therefore
voids the warranty.
CAUTION: Contact of liquid with the LED causes damage and therefore voids the warranty.
CAUTION: Do not spray liquid directly onto the surface of the charging unit and/or of the
handpiece.
CAUTION: The casing of the charging unit and/or the handpiece are not protected against
the entry of liquids.
CAUTION: The charging unit and the handpiece should not be sterilised.
19