6.2
Safety Requirements During Use
! WARNING: Never point the beam of light in the direction of the eyes.
! WARNING: Before each cycle of exposure always make sure that the optical protection has
been fitted onto the handpiece.
! WARNING: Point the beam of light directly onto the material to be polymerised. Do
not subject the gum or other soft tissues to the beam of light (shield these parts suitably if
necessary). the effect of the light should be limited to the oral cavity and in particular to the
sector requiring clinical treatment..
! WARNING: Do not insert metal objects into the handpiece housing in the charging unit
when the device is on.
CAUTION: During the first few seconds of exposure avoid contact of the tip with the
material to be polymerised. Deposits of composite material adhering to and polymerised
to the tip terminal surface lower the amount of light transmitted and will therefore prejudice
subsequent polymerisation operations.
! WARNING: During the intervention on the patient, do not perform any maintenance tasks
on the system.
6.3
Instructions for Use
starlight uno allows performing 2 types of exposures:
• FAST: exposure time of 10 seconds at the maximum light intensity.
• SLOW RISE: exposure time of 20 seconds with a gradual increase of the light intensity
during the first 3 seconds up to the maximum intensity.
FAST exposure selection:
1.
Press the on/off button on the handpiece (Table 1 at page 11 - Ref. D) for a short time.
An acoustic signal will be emitted (1 beep) and the green LED on the handpiece will light
on.
2.
After 10 seconds an acoustic signal will be produced (1 beep) and the green LED on the
handpiece will light off. the FASt cycle has been completed.
SLOW RISE exposure selection:
1.
Keep pressed the on/off button on the handpiece (Table 1 at page 11 - Ref. D) for 2
seconds. An acoustic signal is emitted when the cycle starts and after 2 seconds. The
yellow LED on the handpiece will light on.
2.
After 10 seconds an acoustic signal will be emitted (1 beep).
3.
After 20 seconds an acoustic signal will be emitted (1 beep) and the yellow LED on the
handpiece will light off. the SLOW RISE cycle has been completed.
After the end of the treatment, remove the protective sheath used and place the starlight uno
handpiece back into the charging unit (Figure 5 at page 8 - Ref. B).