INSTALLATION
The device must be installed in a suitable place that is convenient for its use.
WARNING: The place where the device is installed must meet the requirements found in
the
Chapter 4.1 on page 9
4.1
Safety Requirements in the Installation Phase
! WARNING: The electrical system of the premises in which the device is installed and used
must comply with the rules in force and the relevant requirements of electrical safety.
! WARNING: Risk of explosion. The device cannot operate in environments where there are
saturated atmospheres of flammable gases (aesthetic mixtures, oxygen, etc.).
! WARNING: Install the device in a safe place protected from impact or accidental water or
liquid spray.
! WARNING: Do not install the device above or near sources of heat. Arrange during
installation for a suitable circulation of air around the device.
! WARNING: Do not introduce metal objects into the handpiece housing in the charging
unit (Figure 6 at page 10 - Ref. B) when the device is on.
CAUTION: The device can be transported but must be handled with care when moved.
CAUTION: Do not expose the device to direct sunlight or sources of UV light.
CAUTION: Place the device in such a way as to always have the power plug easily
accessible as it is considered to be a disconnection device.
CAUTION: The handpiece is intended for use within the patient environment whereas the
charger unit is not intended for use within the patient environment.
NOTA: the patient environment is defined as an area 1.5 m from the patient (as per IEC 60601-
1, third edition).
! WARNING: The operator must not come simultaneously into contact with the elements
outside the patient environment (charger units and switching power supply adapter) and the
patient. Do not connect other external components to the medical system.
.