Stryker S3 MedSurg Bed with StayPut Frame Mode D'emploi page 8

Masquer les pouces Voir aussi pour S3 MedSurg Bed with StayPut Frame:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

• Do not use the 110V auxiliary power outlet (option) for life sustaining equipment.
• Always lock the control panel when you leave the patient unattended.
• Always lock the patient control panel when the patient's condition requires extra safety measures.
• Do not use bed exit to replace patient monitoring protocol. Bed exit is intended only to aid in the detection of a patient
exiting the product.
• Do not use bed exit with patients who weigh less than 50 lb (23 kg).
• Do not use the scale system reading as a reference for medical treatment. The scale system assists only in the
monitoring of the patient's weight variation.
• Do not set the scale to zero or weigh the patient when a support surface therapy is active. Motion from the support
surface functions may adversely affect the scale system performance.
• Do not use i i B B e e d d Awareness to replace your patient monitoring protocol.
• Do not use i i B B e e d d Awareness as a lock indicator for siderails. i i B B e e d d Awareness only detects the position of the siderails.
C C A A U U T T I I O O N N
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Always make sure that product operation is not affected when you attach equipment to the product.
• Always make sure that all persons and equipment are away from the area below and around the Fowler before you
activate the CPR release. The CPR release is for emergency use only.
• Always follow hospital protocol to determine the use of restraint straps and restraint strap locations. Stryker is not
responsible for the type or use of restraint straps on any Stryker product.
• Do not load the IV pole above the safe working load of 40 lb (18 kg).
• Always raise the siderails when the litter is in its full down position. This prevents the scale system from weighing a
patient inaccurately.
• Always make sure that the siderails are locked before you arm i i B B e e d d Awareness.
• Always make sure that you set the desired product parameters before you arm i i B B e e d d Awareness.
• Do not use accessories that cover the footboard LED light bars.
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the patient from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until it is completely
dry and you have tested for safe operation.
• Always clean V V e e l l c c r r o o
disinfectant for nylon V V e e l l c c r r o o should be determined by the hospital.
• Always wipe down with clean water (or 70% isopropyl alcohol, if using V V i i r r e e x x ® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and
dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause
premature degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.
• Always clean V V e e l l c c r r o o ® after each use. Saturate V V e e l l c c r r o o with disinfectant and allow disinfectant to evaporate. Appropriate
disinfectant for nylon V V e e l l c c r r o o should be determined by the hospital.
• Do not use quaternary disinfectants formulated with glycol ethers.
EN
4
® ®
after each use. Saturate V V e e l l c c r r o o with disinfectant and allow disinfectant to evaporate. Appropriate
3006-209-005 Rev AA.2

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

3002

Table des Matières