Stryker S3 MedSurg Bed with StayPut Frame Mode D'emploi page 62

Masquer les pouces Voir aussi pour S3 MedSurg Bed with StayPut Frame:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

D D i i s s i i n n f f e e c c t t i i n n g g
C C A A U U T T I I O O N N
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the patient from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until it is completely
dry and you have tested for safe operation.
• Always wipe down with clean water (or 70% isopropyl alcohol, if using V V i i r r e e x x ® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and
dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause
premature degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.
• Always clean V V e e l l c c r r o o ® after each use. Saturate V V e e l l c c r r o o with disinfectant and allow disinfectant to evaporate. Appropriate
disinfectant for nylon V V e e l l c c r r o o should be determined by the hospital.
• Do not use quaternary disinfectants formulated with glycol ethers.
The recommended disinfectants for this product's surfaces include the following:
Quaternary (active ingredient - ammonium chloride) that contain less than 10% glycol ether.
Phenolic (active ingredient - o-phenylphenol)
Chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water)
Isopropanol 21.000%
To wipe down the product with disinfectant between uses:
1. Follow the manufacturer's dilution recommendations exactly.
2. Apply the recommended disinfectant solution by spray or pre-soaked wipes.
3. Hand wash all surfaces of the product with the recommended disinfectant.
4. Disinfect all exposed surfaces.
5. Follow the disinfecting solution manufacturer's instructions for appropriate contact time and rinsing requirements.
6. Dry the product thoroughly before returning the product to service.
Avoid over saturation. Do not allow the product to remain wet.
Follow the manufacturer's dilution recommendations for appropriate contact time and rinsing requirements. Follow the
chemical manufacturer's guidelines for proper disinfecting.
EN
58
3006-209-005 Rev AA.2

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

3002

Table des Matières