F F i i g g u u r r e e 3 3 3 3 – – D D o o n n o o t t t t o o u u c c h h b b e e d d
A confirmation notification indicates that setting the scale to zero was successful (Figure 34).
F F i i g g u u r r e e 3 3 4 4 – – Z Z e e r r o o i i n n g g s s u u c c c c e e s s s s f f u u l l
N N o o t t e e
• The B B e e d d Z Z e e r r o o LED on the footboard LED indicator illuminates.
• The scale icon and 0.0 appears on the footboard display.
F F i i g g u u r r e e 3 3 5 5 – – S S c c a a l l e e s s e e t t t t o o z z e e r r o o
N N o o t t e e - - If you first receive an U U n n a a b b l l e e t t o o Z Z e e r r o o - - T T r r y y A A g g a a i i n n notification, the scale attempts to set the scale to zero again for
30 seconds. After three attempts, the scale system locks and an U U n n a a b b l l e e t t o o Z Z e e r r o o notification appears.
M M e e n n u u d d i i s s p p l l a a y y
The S S 3 3 footboard control panel has a menu that displays the menus for S S 3 3 functions and features.