Remove the product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
All fasteners are secure
Apply brake pedal and push on bed to make sure that all casters lock
LED on the footboard and head end siderails blink when brakes are released
Locking steer caster locks and unlocks
Siderails move, latch, and stow
CPR release operable
Foot prop intact and operable
IV pole is intact and operable (option)
Foley bag hooks intact
No cracks or splits in headboard, footboard, or siderail panels
No rips or cracks in mattress cover
All functions on head end siderails operable (includes LEDs)
All functions on footboard operable (includes LEDs)
Scale and bed exit system calibrated
Motion interrupt switches operable
Night light operable
Power cord and plug not frayed or damaged
No cables worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 200 mΩ (milliohms)
Current leakage not more than 300 µA (microamps)
Apply grease to the litter grease points
Motorized drive wheel functions (Z Z o o o o m m option)
Motion release switches operable ( ( Z Z o o o o m m option)
All functions on head end control panel operable (Z Z o o o o m m / battery backup option)
Confirm battery power functions (Z Z o o o o m m / battery backup option)
Power cord slack at the inlet filter (Z Z o o o o m m / battery backup option)
Ground chains are clean, intact, and have at least two links touching the floor
Fowler angle accuracy is 0° - 60°
Fowler holds position at 30° with patient weight
Siderail switches operable ( i i B B e e d d Awareness option)
Center light bar LED and side light LED operable ( i i B B e e d d Awareness option)
Inspect footboard control labels for signs of degradation
3006-209-005 Rev AA.2