• Do not use i i B B e e d d Awareness to replace your patient monitoring protocol.
• Do not use i i B B e e d d Awareness as a lock indicator for siderails. i i B B e e d d Awareness only detects the position of the siderails.
C C A A U U T T I I O O N N
• Always make sure that the siderails are locked before you arm i i B B e e d d Awareness.
• Always make sure that you set the desired product parameters before you arm i i B B e e d d Awareness.
• Do not use accessories that cover the footboard LED light bars.
When enabled, i i B B e e d d Awareness helps to monitor S S 3 3 status and parameter conditions.
To monitor a parameter, place the product to the desired position. You can monitor the low height position, C C h h a a p p e e r r o o n n e e bed
exit, and Fowler 30°+.
When armed, i i B B e e d d Awareness automatically monitors all current siderail positions and the brake.
To arm i i B B e e d d Awareness, press i i B B e e d d (A) ( i i B B e e d d Awareness (option) (page 47)).
N N o o t t e e - - If there is an error in one of the product functions, an error code will appear. i i B B e e d d Awareness will not arm. For more
information about error codes, see the maintenance manual.
After you arm i i B B e e d d Awareness, the LED light bars on the footboard illuminate green and the monitored footboard LED
indicators on the footboard illuminate.
To set an alarm tone for i i B B e e d d Awareness, see Setting an audible i i B B e e d d Awareness alarm (page 46).
To disarm i i B B e e d d Awareness, press i i B B e e d d again (A) ( i i B B e e d d Awareness (option) (page 47)).
N N o o t t e e - - The settings for lockout controls, scale calibration data, bed exit, and i i B B e e d d Awareness are preserved when the
product is unplugged, or during a power failure.
A A c c k k n n o o w w l l e e d d g g i i n n g g i i B B e e d d A A w w a a r r e e n n e e s s s s s s t t a a t t u u s s a a l l e e r r t t s s
If the parameter conditions selected for i i B B e e d d Awareness are changed:
•
LED light bars on the footboard flash amber
•
changed indicator LED on the footboard LED indicator flashes
•
sound alarm is triggered
•
changed parameter status alert is displayed on the display screen
If the low height position changes (Figure 53), return the product to low height (see operator control panel or the footboard
control panel).
If the brakes are released (Figure 54), apply the brake (see Applying or releasing the brakes (page 14)).
EN
48
F F i i g g u u r r e e 5 5 3 3 – – L L o o w w H H e e i i g g h h t t s s t t a a t t u u s s a a l l e e r r t t