• Only use hospital-grade electric equipment consuming 10A or less with the auxiliary power outlet (option). The use of
standard electric equipment may bring the current leakage to a level unacceptable for hospital equipment.
• Do not use the 110V auxiliary power outlet (option) for life sustaining equipment.
The 110 volt auxiliary power outlet is a built-in outlet for peripheral equipment. You can find the outlet under the foot end on
the patient right side of the product ( Product illustration (page 9)).
O O p p e e r r a a t t o o r r c c o o n n t t r r o o l l p p a a n n e e l l ( ( o o u u t t s s i i d d e e s s i i d d e e r r a a i i l l ) )
A
Brake LED
Cardiac chair position
B
3006-209-005 Rev AA.2
F F i i g g u u r r e e 2 2 3 3 – – O O p p t t i i o o n n a a l l a a u u x x i i l l i i a a r r y y o o u u t t l l e e t t
A
B
C
E
D
Flashes amber when you release the brakes. Brake
LED turns off when you apply the brakes.
Press and hold to place the product into the cardiac
chair position
F
H
G
I