CAUTION: Check your system setup by viewing the display on the smartphone Masimo SafetyNet
Alert Application. The Masimo SafetyNet Alert App will provide an indication if there is a problem
with the internet connection.
CAUTION: Avoid placing Masimo SafetyNet Alert near other wireless devices to prevent loss of
performance.
Masimo Sensor
WARNING: Properly apply the Masimo Sensor according to sensor's Directions for Use. Applying
the sensor incorrectly could result in incorrect or no readings.
CAUTION: Avoid using Masimo Sensor under bright light sources and direct sunlight to maintain
the performance of the device.
CAUTION: To maintain Bluetooth connectivity with the Masimo Sensor, ensure that the Home
Medical Hub is within the specified distance and in line-of-sight of the Masimo Sensor. See
Specifications
on page 51.
Cleaning and Service Warnings
WARNING: Do not attempt to remanufacture, recondition or recycle the Masimo Sensor, Masimo
Chip or Home Medical Hub to prevent harm or damage to the system.
WARNING: Always turn off and unplug the Home Medical Hub before cleaning to prevent harm
or damage to the device.
CAUTION: Do not clean Masimo Sensor, Masimo Chip or Home Medical Hub with undiluted
bleach, petroleum-based products, acetone, or other harsh solvents. Clean only with the solutions
specified in this manual to prevent damage to the device.
CAUTION: Do not submerge Masimo Sensor, Masimo Chip or Home Medical Hub in liquid or
attempt to sterilize by any method to prevent damage to the device.
Compliance Warnings
WARNING: Changes or modifications not approved by Masimo can void the user's authority to
operate the equipment.