Accuracy (ARMS*)
Oxygen Level (SpO
)
2
Range
70% to 100%
No Motion [1]
Adults, Pediatrics
Neonates
Motion [2]
Adults, Pediatrics, Neonates
Low perfusion [3] Adults, Pediatrics
Neonates
Pulse Rate (PR)
Range
25 bpm to 240 bpm
No motion
Adults, Pediatrics, Neonates 3 bpm
Motion
Adults, Pediatrics, Neonates 5 bpm
Low Perfusion [4] Adults, Pediatrics, Neonates 3 bpm
* A
accuracy is a statistical calculation of the difference between device measurements and
RMS
reference measurements. Approximately two-thirds of the device measurements fell within +/-
A
of the reference measurements in a controlled study.
RMS
Oxygen Level (SpO2) ARMS Performance Specifications
The tables below provides A
Sensor (Radius PPG) under no motion, with Masimo Technology in a clinical study.
A
Measurement
Values for Radius PPG Sensors
RMS
SpO
Accuracy Range (%)
2
90-100
80-90
70-80
2%
3%
3%
2%
3%
(Accuracy Root Mean Square) values measured using the Masimo
RMS
ARMS (%)
1.14
1.29
1.41
52