Electromagnetic Immunity
Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below. The customer or
the user of the ME Equipment should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Electrostatic
+/- 8 kV contact
discharge (ESD)
+/- 15 kV air
IEC 61000-4-2
Electrical fast
+/- 2 kV for power
transient/ burst
lines
IEC 61000-4-4
+/- 1 kV for input/
output lines
Surge
+/-1 kV line(s) to
IEC 61000-4-5
line(s)
+/- 2 kV line(s) to
earth
Conducted RF
3 Vrms
IEC 61000-4-6
6 Vrms
Compliance Level
+/- 8 kV contact
+/- 15 air
+/- 2 kV for power lines
+/- 1 kV for input/
output lines
+/-1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
3 Vrms
6 Vrms
59
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Performed over 0.15-80 MHz
Performed on the following ISM
(industrial, scientific and medical) bands
of frequency: The bands between 0,15
MHz and 80 MHz are 6,765 MHz to 6,795
MHz; 13,553 MHz to 13,567 MHz;
26,957 MHz to 27,283 MHz; and 40,66
MHz to 40,70 MHz. The amateur radio
bands between 0,15 MHz and 80 MHz are
1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz,
5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz,
10,1 MHz to 10,15 MHz, 14 MHz to 14,2
MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz
to 21,4 MHz, 24,89 MHz to 24,99 MHz,
28,0 MHz to 29,7 MHz and 50,0 MHz to
54,0 MHz