Test
Band
Frequency
(a)
(MHz)
(MHz)
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and
the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-
3.
(a) For some services, only the uplink frequencies are included.
(b) The carrier shall be modulated use a 50% duty cycle square wave signal.
(c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
Recommended Separation Distances
Recommended Separation Distance Between Portable and Mobile RF Communication Equipment
and the ME Equipment
The ME Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the ME Equipment can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the ME Equipment as recommended below, according to the maximum output power of the
communication equipment.
Rated maximum output power of
transmitter (W)
0.01
0.1
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Service (a)
1
10
Modulation
Maximum
(b)
Power
(W)
Separation Distance According to Frequency of
Transmitter (m)
d = 0.6*Sqrt (P)
62
Distance
Immunity
(m)
Test Level
(V/m)
0.06
0.19
0.6
1.9
6