Electromagnetic Immunity
Power frequency
30 A/m
(50 / 60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips on
0% U
power supply input
at 0°, 45°, 90°,
lines
135°, 180°, 225°,
IEC 61000-4-11
270°, and 315°;
0% U
and 70% U
25/30 cycles at 0°
Voltage
0% U
Interruptions on
cycle
power supply input
lines
IEC 61000-4-11
Radiated RF
10 V/m
IEC 61000-4-3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the ME Equipment is used
exceeds the applicable RF compliance level above, the ME Equipment should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the ME Equipment.
1
U
: Rated voltage for the equipment.
T
30 A/m
1
, 0.5 cycle,
0% U
1
, 0.5 cycle, at 0°,
T
T
45°, 90°, 135°, 180°,
225°, 270°, and 315°;
0% U
1 cycle, and 70%
T
1 cycle,
U
25/30 cycles at 0°
T
T
T
, 250/300
0% U
, 250/300 cycle
T
T
10 V/m
Power frequency magnetic fields should
be at levels characteristic of typical
location in a typical hospital environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Performed over 80 MHz to 2.7 GHz