Oxygen Level

The Oxygen Level screen shows the Oxygen Level (SpO
Limit (Time Delay) for triggering the different Alert Events. The triggers are different depending
on the Option selected.
Selecting the "?" icon next to the Option displays additional information about the Option.
Low Oxygen Level and Alert Time Delay Default and Available Settings
For COVID-19 or Other, you have the option to change the default Oxygen Level Low Limits and
Alert Time Delays. Changes to the limit affect when Alert Events are triggered. Default settings can
also be restored if needed.
Alert Event Defaults and Settings (COVID-19)
Alert Type
Low SpO
Default
Level 1 -
93
Caution
Level 2 -
90
Warning
Level 3 -
80
Emergency
Alert Event Defaults (Opioids)
Alert Type
Low SpO
Default
Level 1 -
85
Caution
Level 2 -
80
Warning
80
Level 3 -
Emergency
60
Limit %
Low SpO
2
2
Settings
Max: 95
Min: Warning Limit +1%
Max: Caution Limit –1%
Min: Emergency Limit +1%
Max: Warning Limit – 1%
Min: 80
Limit %
Low SpO
2
Settings
NA
NA
NA
NA
) Low Limit and the time below the Low
2
Limit % User
Time Delay
Default
5 minutes
5 minutes
2 minutes
Limit % User
Time Delay
2
Default
30 seconds
30 seconds
180 seconds
60 seconds
36
Time Delay User
Settings
30 seconds to 5 minutes
in 30 second increments
30 seconds to 5 minutes
in 30 second increments
30 seconds to 2 minutes
in 30 second increments
Time Delay User
Settings
NA
NA
NA
NA