[3] The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing against a
Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and
transmission of greater than 5% for saturations ranging from 70% to 100%.
[4] The Masimo SET Technology has been validated for pulse rate accuracy for the range of 25-240 bpm in
bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater
than 0.02% and transmission of greater than 5% for saturations ranging from 70% to 100%.
*Registered trademark of Fluke Biomedical Corporation, Everett, Washington.
Service and Maintenance
Cleaning
The Home Medical Hub and Masimo Chip are reusable devices. The devices are supplied and
used non-sterile.
The Masimo Chip should be cleaned before and after it has been applied to a user and/or in
accordance with local and governmental regulations to minimize the risk of cross-contamination.
Smart Phone Cleaning
To properly clean the smart phone, refer to the smart phone's Operator's Manual or Directions For
Use.
Home Medical Hub Cleaning
CAUTION: Check the Home Medical Hub for possible cracks or opening before cleaning.
CAUTION: Do not allow liquids to enter the interior of the Home Medical Hub.
The outer surfaces can be cleaned either with a soft cloth dampened with a mild detergent and
warm water solution or they can be wiped down with the following cleaning solutions:
•
70% isopropyl alcohol
•
1:10 bleach to water solution (0.5% sodium hypochlorite)
•
Super Sani-Cloth® Wipes (55% isopropyl alcohol, 0.5% quaternary ammonium
chloride)
•
Windex® (1.5% 2-Butoxyethanol, 1.5% ethylene glycol hexyl ether, 5% isopropyl
alcohol)
65