Compliance with IEC 60601-1-2:2014
Use of X-Smart IQ
avoided because it could result in improper operation. If such use is necessary,
X-Smart IQ
they are operating normally.
X-Smart IQ
peripherals such as antenna cables and external antennas and it should be used at a
minimal distance of 30 cm (12 inches) from any part of the X-Smart IQ
These include cables specified by the manufacturer. Failure to observe this may result
in the degradation of the performance of X-Smart IQ
This equipment does not have Essential Performance (Ref. IEC 80601-2-60).
Guidance and manufacturer's declaration - electromagnetic emissions
®
X-Smart IQ
Handpiece is intended for use in the electromagnetic environment specified below.
The user should ensure that it is used in such an environment.
Emission test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
48/52
®
Handpiece adjacent to, or stacked with other equipment should be
®
Handpiece and the other equipment should be observed to verify that
®
Handpiece is a portable RF communication equipment which includes
Compliance
X-Smart IQ
conductively coupled radio-frequency energy only for
its internal function.
Group 1
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
X-Smart IQ
establishments directly connected to low voltage
power supply network which supplies buildings used
Class B
for domestic purposes.
Professional healthcare facility environment.
Not applicable
-
Not applicable
-
®
Handpiece.
Electromagnetic environment - guidance
®
Handpiece generates and uses
®
Handpiece is suitable for use in
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
®
Handpiece.