Cleaning, Disinfection and Sterilization
7.5.3.1
Approved Sterilization Procedure of the X-Smart IQ
N°
Operation
Packaging
Place the X-Smart IQ
A
Contra-angle in sterile barrier
packages compliant with
EN 868-5.
Sterilization
Dynamic Air Removal
(pre-vacuum) Sterilization
Sterilize at 132°C (270°F) for
4 min.
Dynamic Air Removal
(steam pulse pressure
pulse) Sterilization
Sterilize at 132°C (270°F) for
4 min.
B
Gravity-Displacement
Steam
Sterilize at 121°C (250°F) for
30 min.
Gravity-Displacement
Steam
Sterilize at 134°C for 3 min
(EU Market Only).
Gravity-Displacement
Steam
Sterilize at 132°C (270°F) for
15 min.
Drying / Storage
C
Drying time for all methods is
30 minutes.
32/52
Clean and disinfect before sterilization.
The X-Smart IQ
®
lubricated before sterilization.
Make sure to use only dry sterile items.
Use a sterilization temperature with tolerance in
compliance with ISO 17665.
Comply with your national guidelines, standards and
requirements for cleaning, disinfection and sterilization.
Steam sterilization in accordance with ANSI/
AAMI ST79.
Before resuming operation: wait until the X-Smart IQ
Contra- angle has cooled down and is completely dry.
Store sterile X-Smart IQ
free place at room temperature.
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
®
Remarks
®
Contra-angle should also be
®
Contra-angles in a dry, dust-
Contra-angle
®