Guidance and manufacturer's declaration electromagnetic immunity
®
X-Smart IQ
Handpiece is intended for use in the electromagnetic environment specified below.
The user should assure that it is used in such an environment.
IEC 60601 test
Immunity test
3 Vrms
150 kHz -
80 MHz outside
ISM and
amateur radio
Conducted
bands
RF IEC
61000-4-6
10 Vrms
150 kHz -
80 MHz in ISM
and amateur
radio bands
Radiated RF
10 V/m
IEC
80 MHz to
61000-4-3
2.5 GHz
46/52
Compliance
level
level
10 Vrms
10 Vrms
10 V/m
Electromagnetic environment guidance
Portable and mobile RF communications
equipment should be used at a minimal
distance from any part of X-Smart IQ
Handpiece, including cables. The
recommended separation distance can be
calculated from the equation applicable to the
frequency of the transmitter.
d
d
d
0.35
P
=
80 MHz to 800 MHz
d
0.7
P
=
800 MHz to 2.7 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
b
meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
a
survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
®
0.35
P
=
0.35
P
=
b