± 1 kV differential
Surge IEC
mode
61000-4-5
± 2 kV common mode
< 5% U
(> 95% dip in U
for 0.5 cycle
Voltage dips,
< 40% U
short
(60% dip in U
interruptions
for 5 cycles
and voltage
variations on
power supply
70% U
input lines IEC
(30% dip in U
61000-4-11
for 25 cycles
< 5% U
(> 95% dip in U
5 s
Power
frequency (50 /
60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
± 1 kV differential
mode
± 2 kV common mode
0% U
(> 95% dip in U
T
)
for 0.5 cycle
T
(at 0, 45, 90, 135, 180,
225 and 315° phase
angle)
0% U
T
)
(> 95% dip in U
T
for 1 cycle
(at 0 phase angle)
70% U
T
)
(30% dip in U
T
for 25/30 cycles
(at 0 phase angle)
0% U
T
) for
(> 95% dip in U
T
for 250/300 cycles
(at 0 phase angle)
30 A/m
Mains power quality should
be that of a typical
commercial or hospital
environment.
T
)
T
Mains power quality should
be that of a typical
commercial or hospital
environment.
T
)
If the user of X-Smart IQ
T
Handpiece requires
continued operation during
power mains interruptions, it
is recommended that X-
T
)
Smart IQ
T
powered from an
uninterruptible power supply
or a battery.
T
)
T
Power frequency magnetic
fields should be at levels
characteristic of a location in
a typical commercial or
hospital environment.
®
®
Handpiece be
45/52