Pre-Treatment Of The X-Smart Iq ® Contra-Angle - Denstply Sirona X-Smart IQ Protocole D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 54
Cleaning, Disinfection and Sterilization
X-Smart IQ
Pre-Clean/
Method
Wipe down
®
X-Smart IQ
Contra-angle sterilization and drying time with Gravity-Displacement Steam
Gravity –
Displacement
Yes
Steam
Gravity –
Displacement
Yes
Steam (EU
Market Only)
Gravity –
Displacement
Yes
Steam
7.5.1
Pre-Treatment of the X-Smart IQ
Detach the used endodontic file and separate the X-Smart IQ
the X-Smart IQ
No special transport container is required for the X-Smart IQ
Transportation is not subject to any special requirements.
Rinse the X-Smart IQ
coarse soiling if necessary.
Manual pre-cleaning is required for an efficient cleaning procedure, used in conjunction
with machine-based reprocessing or sterilization.
30/52
®
Contra-angle cleaning and disinfection parameters
Lubricate (for
complete
(Minimum):
instructions,
Temp / Time
see
Section 7.4)
Sterilize at
Before
121°C (250°F) /
30 min
Sterilize at
Before
134°C / 10 min
Sterilize at
Before
132°C (270°F) /
15 min
®
Handpiece, prior to any manual pre-treatment.
®
Contra-angle under running water in order to remove any
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
Drying Time
Drying time is
30 min
Drying time is
30 min
Drying time is
30 min
®
Contra-angle
®
Contra-angle from
®
Contra-angle.
DFU Section/
Page
For complete
instructions, see
Section
7.5.3
For complete
instructions, see
Section
7.5.3
For complete
instructions, see
Section
7.5.3

Publicité

Table des Matières
loading

Table des Matières