Cleaning, Disinfection and Sterilization
•
Do not place the X-Smart IQ
disinfector.
•
Wear protective clothing during all cleaning, disinfection and sterilization
processes.
•
Do not apply fast sterilization or sterilization procedures with non-packaged
accessories. In addition, do not use hot-air sterilization, radiation sterilization or
sterilization with formaldehyde, ethylene oxide or plasma.
•
Never place the components of the kit (listed in table below) in a washer-
disinfector, autoclave or ultrasonic bath.
•
If you use a disinfecting spray, never spray the devices and accessories directly.
Instead, spray a paper towel or use a pre-moistened disinfectant towelette (wipe).
•
Only use surface disinfectants that are certified by officially recognized institutes,
and are chlorine-free and aldehyde-free.
•
In addition to using the X-Smart IQ
X-Smart IQ
wipe-down disinfection ensure that the charging cable is not connected and that
the charging port is closed.
•
Do not use CaviCide1™ or CaviWipes1™ on any items shown or listed in table
below.
See Endo IQ
For wipe-down disinfection of components proceed as follows:
N°
After any pre-cleaning, lightly soak a paper tissue or soft cloth in a mild, aldehyde-free
disinfection and cleaning solution (bactericidal and fungicidal) and use it to clean /
A
disinfect all components.
Dentsply Sirona recommends the use of Dentsply VoloWipes™.
Thoroughly disinfect surfaces with a lightly soaked wipe (min. 30 seconds), making sure
B
the entire surface is wet. Allow contact time > 30 minutes. If large surfaces have to be
treated, use more than one wipe where necessary.
26/52
®
Handpiece in an autoclave unit or washer-
®
Handpiece regularly. When subjecting X-Smart IQ
®
App User Manual -
6 Step by Step
®
Handpiece Sleeve, clean and disinfect the
Instructions.
Action
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
®
Handpiece to