Denstply Sirona X-Smart IQ Protocole D'utilisation page 36

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 54
®
X-Smart IQ
Handpiece Error Codes
State
36/52
Failure
Wait for a few minutes and press the multifunction
Motor
overheating
button to reactivate X-Smart IQ
Motor hardware
Reset the device by holding the multifunction button
failure
for 10 seconds (see
Return the X-Smart IQ
Battery failure
center.
General
Reset the device by holding the multifunction button
hardware failure
for 10 seconds (see
®
X-Smart IQ
Handpiece
Reset the device by holding the multifunction button
START/STOP
for 10 seconds (see
button sensor is
not working
properly
Bluetooth radio
Reset the device by holding the multifunction button
module failure
for 10 seconds (see
Lubricate the X-Smart IQ
Sterilization of the X-Smart IQ
(Autoclaving)) and perform the calibration again.
Calibration failed
If the error persists, try to replace the X-Smart IQ
Contra-angle and perform the calibration again.
Solution
6.5.1 General
6.5.1 General
6.5.1 General
6.5.1 General
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
®
Handpiece.
Overview).
®
Handpiece to your service
Overview).
Overview).
Overview).
®
Contra-angle (see
®
Contra-angle
7.5.3
®

Publicité

Table des Matières
loading