NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobiles radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which X-Smart IQ
compliance level above, X-Smart IQ
abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating X-Smart IQ
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications equipment
®
X-Smart IQ
Handpiece is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The user can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and X-Smart IQ
output power of the communications equipment.
Rated
maximum
output power of
150 kHz to 80 MHz
transmitter
(W)
0.01
0.04
0.1
0.13
1
0.4
10
1.3
100
4.0
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
®
Handpiece should be observed to verity normal operation. If
®
Handpiece or the ME EQUIPMENT in its vicinity.
and X-Smart IQ
®
Handpiece as recommended below, according to the maximum
Separation distance according to frequency of transmitter
d
0.35
P
=
0.04
0.11
0.35
1.1
3.5
®
Handpiece is used exceeds the applicable RF
®
Handpiece
(m)
80 MHz to 800 MHz
d
0.35
P
=
800 MHz to 2.7 GHz
d
0.7
P
=
0.07
0.22
0.7
2.2
7.0