9.2
GUIDE TO PROBLEMS
The table below represents a guide to the potential defects of the light.
If you cannot resolve the problems, call technical support
.
Effect
The light does not
switch on
The light will flicker
The light does not
adjust light intensity
The light intensity is
considerably reduced
Stains have appeared
on the parabolas or
the reflective layer has
come away.
The light does not stay
balanced and tends to
lower
The lamp does not
control
64
Cause
Action (Service Engineer - SE)
Power supply not
Check the power supply is inserted and
inserted or inserted
the combined unit is on.
incorrectly.
Interference with
Switch off the electro-surgical scalpels
electro-surgical
and check the permanence of the effect.
scalpels or high
energy tools
Command on joystick
To switch on and off the light, press and
applied incorrectly.
release the joystick lever using the right
or left side.
Interference with
Switch off the electro-surgical scalpels
electro-surgical
and check the permanence of the effect.
scalpels or high
energy tools.
Command on joystick
Use the command correctly as described
applied incorrectly.
in this manual.
Switch off the electro-surgical scalpels
Interference with
and check the permanence of the
electro-surgical
effect.
scalpels or high
energy tools.
Parabolas or
Clean the parabolas and the secondary
lenses.
secondary lens dirty.
Check you have maximum adjustment
Use of wrong
with the command.
procedures.
Clean the surfaces with specific "Faro
Use of non-approved
Perflex" product. Clean the surfaces with
products.
isopropyl alcohol.
To restore the surfaces, you need to get
service to replace the parabolas.
Excessive load on
Remove the excess loads.
head (small mirrors,
cameras, etc..).
Activate the function; see 4.1.4
Synchro function off
Resp
User
User
User
User
User
User
User
User
User
User
User