Wall Mounting - Faro ALYA Mode D'emploi

Lampada dentale a led
Masquer les pouces Voir aussi pour ALYA:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

WALL MOUNTING

LIGHT MOUNTING WALL MOUNTING
1. Plug
2. Screw
3. Wall mounting
4. Expanders
5. Terminal board
6. Terminal board cover
7. Cover
8. Screw
8. Washer
8. Cable gland
40°
50°
NB1. The device has to be installed by specialized technicians only
NB2. Always cut off the power supply at the installation site prior to installation
NB3. Before proceeding with the assembly, ensure the wall can bear the application.
The authorized construction materials of the wall are concrete and natural stone. Use the supplied
fittings or equivalent ones.
NB4. Maximum applicable load: 70 kg
NB5. Installation is allowed only in spaces provided with electrical plant compliant with the
national legislation in force as regards spaces for medical use
NB6. The lamp without transformer must be powered with low-voltage (12-24Vac o 17-33Vdc)
using a safety transformer (according to IEC/EN 60601-1) with thermal protection or at least
one suitable fuse (T500mAL250V~).
The resulting medical system has to be declared by the installer as being compliant with IEC/
EN 60601-1
52