7 Setup methods according to DIN EN ISO 17664 (cleaning, disinfection, maintenance and sterilization)
7.1 Preparation at the site of use
WARNING
Hazard from nonsterile products.
There is a risk of infection from contaminated medical devices.
 Take suitable personal protective measures.
 Remove all residual cement, composite or blood without delay.
 Recondition the medical device as soon as possible after treatment.
 Remove the cutter or grinder from the medical device.
 The medical device must be dry when transported for reconditioning.
 Do not place it in a solution or similar.
7.2 Cleaning
CAUTION
Malfunctions from cleaning in the ultrasonic unit.
Defects in the product.
 Only clean manually or in a thermodisinfector.