Guide To Problems - Faro SK07 Mode D'emploi

Masquer les pouces Voir aussi pour SK07:
Table des Matières

Publicité

GUIDE TO PROBLEMS

ERRORS DURING THE CYCLES
CODE
MESSAGE
ER001
Safety Cut Off
ER002
Top-up
ER003
Pressure balance
ER004
No Ready
ER005
Timeout 1
ER006
Vacuum
ER007
Timeout 2
ER008
chamber T High
ER009
chamber T Low
ER010
High Pressure
ER011
Low Pressure
ER012
High theoretic T
ER013
Low theoretic T
ER014
Delta Temperature
ER015 Vacuum Test: vacuum vacuum threshold not reached during vacuum leak test
ER016
EN 13060
ER017
EN 13060
ER018
EN 13060
ER019
Door open
ER020 Timeout water charge failed to fi ll the charge tank
ER021
Emptying water tank
MACHINE FAILURES
CODE
MESSAGE
AL001
Chamber Probe Fault
AL002
Chamber Probe Fault
AL003
Band Probe Fault
AL004
Band Probe Fault
AL005
Timeout water charge
AL006
Pressure Trasducer
AL007
Door Micro
AL008
Safety Micro
AL009
Blackout
AL010
Safety Pressure
AL011
Corrupted Memory
AL012
Safety Micro
AL013
PUSH Error
AL014
POP Error
AL015
Reset Pressed
AL016
Safety Micro-switch
Machine release code:
Press the keys "
", "+" and "-" simultaneously, and wait for the end of the procedure. If the
machine has broken down, contact Faro.
DESCRIPTION
the safety temperature of the chamber or band has been exceeded
the maximum number of top-ups permitted has been exceeded
the pressure balance has not been reached
the ready temperature has not been reached
failure to reach fi nal pressure during thermodynamic discharge
the second vacuum threshold has not been reached
the sterilization conditions have not been reached
chamber temperature is above the limit allowed
chamber temperature is below the limit allowed
the pressure is above the limit allowed
the pressure is below the limit allowed
the theoretical temperature is above the limit allowed
temperature below the limit allowed
the difference between chamber temperature and theoretic
temperature is beyond the limit allowed
parameters not respected in fi rst phase of vacuum leak test
parameters not respected in second phase of vacuum leak test
failure to observe temperature limits during vacuum test
Door not closed
failed to empty the discahrge tank
DESCRIPTION
chamber probe open
chamber probe short-circuited
band probe open
band probe short-circuited
exceeding the maximum time for charging water
pressure transducer fault
unforeseen door micro contact opening
unforeseen door safety micro contact opening
blackout occurred
pressure above safety limit
memory corrupted
Door safety off
Software error
Software error
Reset button pressed
Door safety on
41 41