FLOOR MOUNTING
The device has to be installed by specialized technicians only
NB2. Always cut off the power supply prior to installation
NB3. Before proceeding with the assembly, ensure the floor can bear the application.
The authorized construction materials of the wall are concrete and natural stone. Use the sup-
plied fitting or equivalent ones.
NB4. Maximum applicable load: 70 kg
NB5. Installation is allowed only in spaces provided with electrical plant compliant with the
national legislation in force as regards spaces for medical use
NB6. The lamp without transformer must be powered with low-voltage (12-24Vac o 17-33Vdc)
using a safety transformer (according to IEC/EN 60601-1) with thermal protection or at least
one suitable fuse (T500mAL250V~). The resulting medical system has to be declared by the
installer as being compliant with IEC/EN 60601-1
A. After having fixed as reference the centre of the
chair "a", install at a distance 650mm and 150mm
according to the directions given in figure"C "
- Always cut off the power supply at the installation site prior any operation
- A fter having fixed as reference point (a) the chair centre (see fig.C), drill four d.14 bores in
the floor, so that they match the bores on the support (13).
- P repare the floor support (13) by passing through the washer (7)and screw (6), then screw
the expanders (12) on the screws (6) a few turns, pass the power cable through the cable gland
(11).
- I nsert the expanders (12) in the d.14 bores previously carried out and block with the
special hex spanner (installation accessory) the 6 screws; do not choke the cable between the
support (13) and the floor.
- A pply four plugs (5) on the floor support (13) bores.
- U ntighten the screws (8) and remove the cover plate (9) Connect the power supply
cable to the terminal board (10).
- F ix the column (4) to the floor support (13); while securing it, ensure the column is perpen-
dicular. Secure with the three screws (3) the bushing (2) to the column (4); ensure the holes of
the bushing (2) match the screws on the column (4).
- C onnect the light cables to the terminal board (10).
- F it the cover plate (9) to the floor support (13) with two screws (8).
WIRING DIAGRAM
Floor mounting
Light cables
Terminal board (5)
a
fig. C
Mains cable earthing connection
Mains power supply
12-24 Vac
17-33 Vdc
55 55