SPECIFICATIONS

Power Supply Specifications
Power supply
Operating Condition
Voltage Range
100-240
Frequency
50/60
Current
0.6
Environmental Operating Conditions
Environmental Condition
Operating Condition
Altitude
3000 (max)
16 – 40
Operating Temperature
-10 to 50
Storage Temperature
Operating Relative Humidity
20-80 (non-condensing)
Voltage Range
12
Frequency
DC
Current
1.2
Installation/Over Voltage
Category II
Pollution Degree
2
4
INSTRUCTIONS FOR USE – INITIAL SET-UP
Choose Location
Place 3M™ Attest™ Auto-reader 490 on a firm level surface away from sunlight and
Units
incandescent light. To save counter space, units may be stacked two high.
Volts AC
Plug In Unit
Hertz
Connect the power cord to the power supply. Connect the power supply to the 3M™
Amps
Attest™ Auto-reader 490 and plug the power cord into an AC grounded outlet.
Units
WARNING:
Meters
• Do not place the instrument in environment exposed to sunlight or strong
incandescent light.
ºC
• Do not place instrument close to any device that emits a strong
ºC
electro-magnetic field.
%
• Do not use on a vibrating surface.
Volts DC
Wait 30 Minutes Before Using
Once plugged in, allow a 30 minute warm-up period before placing biological indicators
Amps
into the incubation wells. The C1 caution code will disappear when proper incubation
temperature is reached.
Special Considerations
It is recommended that the unit be left on to eliminate warm-up periods. The unit
may be turned off when it is anticipated that it will not be used for an extended period.
WARNING: Use only the power supply specified for this product and certified
for the country of use.