3. Power-up the terminal and configure for use with a PowerDeck floor scale. Navigate in setup to
4. Remove power from the terminal and connect the PowerDeck platform via the provided M12
5. Power-up the terminal.
6. Respond the "Use factory calibration?" prompt with a Yes or No. If No is selected, the restore
7. If Yes is selected, the terminal will read the stored System Configuration Data from the load
8. If data restore is successful, the terminal will prompt for a GEO code entry. If there was a GEO
9. Once the GEO code is confirmed, the terminal completes the configuration process.
10. Once the scale data restore is complete, the platform should be levelled using the utility
2-46
METTLER TOLEDO IND570 Terminal Installation Manual
Scale > Type > Application and select Floor. Program the # of Load Cells to 4 (four).
connector.
process is cancelled and normal power-up continues.
cells and load them into the terminal.
value in the System Configuration Data, then the terminal pre-populates the display with this
value. Enter the correct GEO value for the installation location and press OK to accept. GEO
values can be found in Appendix D, GEO Codes.
provided in setup at Maintenance > Run > Diagnostics > Scale > Leveling Guidance. Refer to
the User's Guide Chapter 5, section 5.8 for details.
Figure 2-90: Factory Calibration Prompt
30205319
|
11
|