1. Make sure that the un-terminated end of the POWERCELL cable is cut cleanly.
2. Mark the exterior braided armor 10" (255 mm) from the end of the cable.
3. Use metal shears to cut along the exterior braided armor, taking care not to cut into the cable
4. Trim the exterior braided armor back to the cut mark, and remove stray wires from the cut.
5. Mark the cable jacket 1" (25 mm) cm from the cut end of the exterior braided armor.
6. Using a razor knife, cut carefully around the cable jacket, without cutting into the inner braided
7. Use the razor knife to cut along the cable jacket, so that it can be peeled off the wires. Again,
8. Peel the cut jacket off the inner braided shield.
9. Mark the inner braided shield 1" (25 mm) from the cut end of the cable jacket.
10. Carefully cut around the inner shield at the cut mark, without cutting into its contents. Remove
11. Unscrew the outer sealing portion of the POWERCELL cable gland from the terminal, and
12. Use two wrenches to remove the nut holding the body of the POWERCELL cable gland to the
13. Slide the outer sealing part of the gland down the prepared cable, keeping the outer drain wire
30205319
|
11
|
07/2019
Letter
C
Length of black and green wires and internal drain wire
D
Length of red, blue and white wires
E
Length of outer shield drain wire from end of cable jacket
F
Length of insulation to strip from wires
jacket or the outer drain wire.
shield.
take care to cut only the cable jacket, and not its contents: push the knife deep enough into the
cable jacket until it is just possible to feel the tip of the blade drag on the inner braided shield.
the cut portion of the inner shield from the cable.
disassemble the cable clamp. Set the clamps and screws aside in a safe place.
Figure 2-40: Outer Sealing Gland and Clamp
terminal enclosure, and remove the gland.
outside the body. Position the gland against the cut end of the exterior braided armor. Note the
orientation of the gland – the clamping features should be toward the exterior braided armor.
Description
METTLER TOLEDO IND570 Terminal Installation Manual
Length
70 mm (2.75 in)
229 mm (9 in)
216 mm (8.5 in)
5 mm (0.2 in)
2-25