Mettler Toledo IND570 Manuel D'installation page 81

Masquer les pouces Voir aussi pour IND570:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 330
Some examples of connecting external equipment are shown in Figure 2-92.
2.4.5.12.1.
RS-485 Transmission Line Termination
The RS-485 network should include a terminating resistor, installed between the two lines at or on
the last node. The terminating resistor should match the characteristic impedance of the
transmission line, approximately 120 ohms. This terminating resistor is required when connecting
ARM100 modules to the port.
2-48
METTLER TOLEDO IND570 Terminal Installation Manual
NOTES
1. USE ONLY SHIELDED CABLE
2. MAXIMUM CABLE LENGTH:
1000 FEET (304 METERS)
3. WIRE SIZE:
MAX. 18 AWG (0.823 mm2)
MIN. 24 AWG (0.205mm2)
Figure 2-92: Sample COM1 Connections
30205319
|
11
|