Scale Types & Update Rates Analog load cells, up to twelve 350-ohm load cells (eight for IND570xx) (2 or 3 mV/V).
Analog Load Cell Excitation
Voltage
Minimum Sensitivity
Weight Display
Interface and Function
Update Rates (max) –
Analog terminal versions
Interface and Function
Update Rates (max) – IDNet
terminal versions
Interface and Function
Update Rates (max) –
SICSpro terminal versions
30205319
|
11
|
07/2019
IND570 Specifications
Update rate > 366 Hz.
or
IDNet High-Precision Line (+12V versions only, including T-Brick cell, M-Cell, Point-ADC).
Update rate determined by IDNet base.
or
SICSpro High-precision platforms that include Advanced Setup Mode. Refer to section 1.8.1.3
for details. Update rate approximately 90 Hz.
or
Up to 14 POWERCELL load cells for IND570, and up to 6 POWERCELL load cells for
IND570xx, or single PowerDeck platform. Update rate selectable up to 50 Hz.
10 VDC
0.1 microvolts
Displayed resolution of 100,000 counts for analog load cell scales.
Display resolution for high-precision IDNet and SICSpro bases is determined by the specific
base used.
Weight display: 10 Hz
Internal discrete I/O: 50 Hz
External discrete I/O (ARM100): 25 Hz
PLC cyclic data: 25 Hz
SICS continuous (SIR): 20 Hz
MT Continuous Output: 20 Hz
Continuous Template (serial): 20 Hz (19.2-115.2Kbaud), 14 Hz (9600 baud)
Continuous Template (Eprint): 10 Hz
Weight display: 10 Hz
Internal discrete I/O: 20 Hz
External discrete I/O (ARM100): 20 Hz
PLC cyclic data: 20 Hz
SICS continuous (SIR): 20 Hz
MT Continuous Output: 20 Hz
Continuous Template (serial): 20 Hz (19.2-115.2Kbaud), 14 Hz (9600 baud)
Continuous Template (Eprint): 10 Hz
Weight display: 10 Hz
Internal discrete I/O: 50 Hz
External discrete I/O (ARM100): 25 Hz
PLC cyclic data: 25 Hz
SICS continuous (SIR): 20 Hz
MT Continuous Output: 20 Hz
Continuous Template (serial): 20 Hz (19.2-115.2Kbaud), 14 Hz (9600 baud)
Continuous Template (Eprint): 10 Hz
METTLER TOLEDO IND570 Terminal Installation Manual
1-5