Metrology Line
2.7.2.
Capacity Label Instructions
To further ensure that scale capacity and increment display requirements can be met during
installation, a blue capacity label is included with the terminal. This label can be completed and
adhered to the front of the terminal overlay as shown in Figure 2-125. To prevent any potential
conflicts, when using the capacity label, the metrology line on the display should be disabled in
setup.
The capacity label (shown in Figure 2-124) provides space for the Max, min, and e information for
each range or interval for which the scale is programmed. If only one or two ranges are used, the
unused portion of the label may be cut off with scissors. Written information must be legible and a
minimum of 2mm or 0.08 in. in height. A permanent marker should be used for this information.
Clean any oil or other contaminants from the area of the overlay shown in Figure 2-125 where the
capacity label will be added. Peel the backing from the label and adhere it to the overlay in the
location shown in Figure 2-125, or another location acceptable to the local regulations.
30205319
|
11
|
07/2019
Figure 2-123: Metrology Line on IND570 Display
Figure 2-124: Preparing the Capacity Label
METTLER TOLEDO IND570 Terminal Installation Manual
Cut at dotted line to remove
unused ranges/intervals
2-71