Figure 2-60: POWERCELL Wires Connected in Terminal
2.4.5.7.
POWERCELL PDX Vehicle System Cabling, Panel-Mounted Enclosure
Ground and shield terminations are a critical part of the POWERCELL PDX system's immunity to
noise and electrical surges. Prepare and install the terminal end of the POWERCELL PDX home run
cable as follows.
If the IND570 will be installed into a control cabinet or panel where an entry gland is not used,
prepare the cable as follows.
1. Cut and remove an appropriate length of the outer braided shield, the insulation jacket, the
inner braided shield and insulation as shown in Figure 2-61 and Table 2-7.
Figure 2-61: POWERCELL Cable Preparation – Panel
30205319
|
11
|
07/2019
METTLER TOLEDO IND570 Terminal Installation Manual
2-33