Télécharger Imprimer la page

Stimulator Programmable Characteristics - Boston Scientific Vercise PC Ver DB 1140 Mode D'emploi

Publicité

Vercise™ PC Implantable Pulse Generator DFU

Stimulator Programmable Characteristics

The stimulation parameters are independent for the two DBS Leads such that stimulation of
two different brain targets can have different amplitudes, pulse widths, stimulation rates, and
contact configurations. The two DBS Leads can also have differing monopolar and multipolar
configurations; however, a given DBS Lead contact cannot be programmed as both monopolar and
multipolar. The programmable parameter ranges for the Stimulator are shown in Table 2.
Note: Some frequency combination may not be used. See the Programming Manual for
specific frequency combinations.
Table 2: Stimulator Function
Parameter
1
Amplitude
2, 3
Rate
Pulse Width
Cycle (On/Off)
Stim Ramp On
Contact Connections
Independent Areas of Stim (4
Programs with 4 Areas per Program)
The programmable coverage for each individual contact is limited to 12.7 mA.
1
A programming interlock is enforced to limit the total output current to 20 mA
or less per coverage area. For example, a maximum current output of 12.7 mA
on one contact would limit the total summed current output on the remaining
contacts to 7.3 mA within one coverage area.
The rate is limited to a sum of 255 Hz for a given port.
2
The use of higher frequencies than the ones that are established (250 Hz) is
3
the sole responsibility of the user.
The use of lower pulse widths than established (60-450 μsec) is the sole
4
responsibility of the user.
Vercise™ PC Implantable Pulse Generator DFU
90960114-02 12 of 381
4
Range
0.1 – 20 mA
2 – 255 Hz
20 – 450 µs
1 s – 90 min
1 – 10 s
16
16

Publicité

loading

Ce manuel est également adapté pour:

Nm 3138 55Sc 4252Sc 4254