3. Visually check to ensure that the DBS Lead electrodes are aligned with the DBS Extension
contacts. If they are not aligned, continue to grip the DBS Lead next to the Retention Sleeve
and push to advance the electrodes into alignment with the DBS Extension contacts. If
necessary, back out the lead slightly and then advance the electrodes into alignment again,
until proper alignment can be confirmed.
Note: Be sure to fully insert the DBS Lead into the connector so that the retention sleeve
is located under the setscrew.
4. Do not tighten the Setscrew at this time.
5. Repeat steps 1 through 3 to connect the second DBS Lead to the second DBS Extension.
6. Test the impedance of the connection to ensure that you have properly aligned the DBS Lead
within the DBS Extension connector. See "Intraoperative Testing" in the appropriate lead
implant DFU as listed on your DBS Reference Guide.

Assembling the Tunneling Tool

A Tunneling Tool and Straw are provided to facilitate tunneling of the DBS Extension.
1. Attach the Tunneling Tool Handle to the Shaft by turning the locking mechanism clockwise.
a. Push the locking mechanism at the base of the Tool Handle onto the Shaft.
b. Grasping the Tool Handle and the Tip of the Tunneling Tool, rotate the Shaft back and forth
until the handle seats onto the Shaft.
c. While firmly grasping the Tip of the Tunneling Tool to hold the Shaft stationery, turn the
locking mechanism clockwise until secure.
Correct
Tool Handle
Vercise™ PC Implantable Pulse Generator DFU
Stimulator Implantation
Incorrect
Shaft
90960114-02 3 of 381