Stimulator Programmable Characteristics - Boston Scientific Vercise Gevia Mode D'emploi

Masquer les pouces Voir aussi pour Vercise Gevia:
Table des Matières

Publicité

Stimulator Programmable Characteristics

The stimulation parameters are independent for the two DBS Leads such that stimulation of
two different brain targets can have different amplitudes, pulse widths, stimulation rates, and
contact configurations. The two DBS Leads can also have differing monopolar and multipolar
configurations. The programmable parameter ranges for the Stimulator are shown in Table 2.
Note: Some frequency combinations may not be used. Refer to the Programming
Manual for more information on programming multiple frequencies.
Table 2: Stimulator Function
Parameter
1
Amplitude
2, 3
Rate
4
Pulse Width
Cycle (On/Off)
Stim Ramp On
Contact Connections
Independent Areas of Stim (4 Programs
with 4 Areas per Program)
The programmable coverage for each individual contact is limited to 12.7 mA.
1
A programming interlock is enforced to limit the total output current to 20 mA
or less per coverage area. For example, a maximum current output of 12.7 mA
on one contact would limit the total summed current output on the remaining
contacts to 7.3 mA within one coverage area.
The rate is limited to a sum of 255 Hz for a given port.
2
The use of higher frequencies than the ones that are established (250 Hz) is
3
the sole responsibility of the user.
The use of lower pulse widths than established (60-450 μsec) is the sole
4
responsibility of the user.
Range
0.1 – 20 mA
2 – 255 Hz
20 – 450 µs
1 s – 90 min
1 – 10 s
16
16
Vercise Gevia™ 16 Contact IPG DFU
The Vercise Gevia Stimulator
91168761-02 Page 13 of 419

Publicité

Table des Matières
loading

Table des Matières