Vercise™ PC Implantable Pulse Generator DFU
2. Verify that the retention sleeve on the DBS Extension is located directly under the Setscrew in
the Stimulator header.
Note: The retention sleeve is easily distinguishable from the contacts by its length (see
below).
Contacts Retention
3. Check impedances to verify connections before tightening the Setscrew.
a. Place the Stimulator partially in the subcutaneous pocket.
b. Test impedances using the Remote Control or Clinician Programmer.
4. Pass the Torque Wrench through the slit in the septum located on the side of the Stimulator
header.
5. Tighten the Setscrew in the Stimulator until the Torque Wrench clicks, indicating the Setscrew
is fully secured.
Note: To tighten the Setscrew, rotate the Torque Wrench clockwise. To loosen the
Setscrew, rotate the Torque Wrench counterclockwise.
Vercise™ PC Implantable Pulse Generator DFU
90960114-02 6 of 381
Proximal End of the DBS Lead
Sleeve