Disconnect bipolar scissors from connecting cable and clean thoroughly under running water or
with an aldehyde-free disinfectant solution directly after use (within a maximum of 2 h) to
prevent particles and dried secretions from sticking to the instruments. Blood and tissue
residue should not be allowed to dry on the instruments.
Due to the better results in terms of efficacy and repeatability, machine cleaning should be
preferred to manual cleaning.
NOTICE
Risk of damaging the bipolar scissors!
Do not immerse into hydrogen peroxide (H
To avoid scratching of the electrodes and the insulation, do not use abrasive cleaning
tools such as metal brushes or steel wool!
In order to avoid breaking, do not apply mechanical loads to the ceramic parts!
5.3

Cleaning and Disinfection

5.3.1

Machine Reprocessing

Manual precleaning
Immerse instruments into cold water for 5 min.
Clean the instruments under cold mains water using a soft brush until all visible
contaminations have been removed.
Cleaning
Place instruments into a tray basket on the insert module or on the inserts of the MIS unit and
start the cleaning process.
1 min pre-rinse with cold water
Emptying
3 min pre-rinse with cold water
Emptying
5 min washing at 55°C (131°F) with 0.5% alkaline
or at 45°C (113°F) with enzymatic cleaning agent
Emptying
3 min neutralization with warm mains water (>40°C / >104°F) and neutralizer
Emptying
2 min intermediate rinsing with warm mains water (>40°C / >104°F)
Emptying
28