Stryker GoBed II MedSurg Bed FL28C Guide D'utilisation page 42

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Removable I.V. Pole, 1/2" - FDTSH
The removable I . V . pole FDTSH is a 1/2" diameter I . V . pole with an adjustable height .
WARNING
The scale and bed exit systems must be adjusted if this accessory is added when either system is in operation . Refer to
the "Add / Remove Equipment" procedure
INSTALLATION
Tools Required:
• None
Procedure:
1 .
Insert the I . V . pole into one of the sockets located on the four corners of the bed and either side of the seat section .
OPERATION
CAUTION
Do not use the I . V . pole as a push/pull device .
1 .
To adjust the height of the pole:
• Turn the locking sleeve counterclockwise and raise the telescoping portion
of the pole to the desired height .
• Tighten the locking sleeve .
2 .
Hang the I . V . bags .
WARNING
The weight of the I . V . bags must not exceed 11 lb (5 kg) per hook .
Return To Table of Contents
1-40
(page
1-21) .
280C-009-001 REV B
Tightening Sleeve
www.stryker.com

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Gobed ii lit medsurg fl28c

Table des Matières