Beds require an effective maintenance program, we recommend checking these items annually . Use this sheet for your
records and keep on file .
CHECKLIST
_____ All fasteners secure (reference all assembly drawings) .
_____ The head and foot section molded shells are properly attached to the section structures .
_____ Inspect the oil-impregnated bronze shoulder spacers (found at bed hinge points) for excessive wear . Do not
lubricate these spacers . Replace as needed .
_____ Check the grease present on the components detailed in the "Lubrication Requirements" section of the Maintenance
Manual . Lubricate as needed . Lubricate them at least every two years .
_____ On both sides of the bed, depress fully down the side of the pedal identified with a red sticker and ensure that
the brakes are applied and the bed immobilized . Toggle the pedal to neutral and ensure the brakes are
released .
_____ On both sides of the bed, depress fully down the side of the pedal identified with a green sticker and ensure
that the fifth steer wheel is engaged . Toggle the pedal to neutral and ensure that the fifth wheel disengages .
_____ Siderails move, latch and stow properly .
_____ All controls on the foot end panel working properly, including LEDs .
_____ Calibrate the scale system (see the ¨Scale System Calibration¨ section in the Maintenance Manual) .
_____ All siderail controls working properly (see Head Siderail Function Guide) .
_____ Ensure that the optional nurse call alarm sounds in the nurse station .
_____ Ensure that the optional Communications Package controls operate properly .
_____ Verify the CPR emergency release using both CPR release handles: Completely raise the Fowler and, using
the CPR handle, lower the Fowler gradually to a flat position by pulling, holding and releasing the handle
several times . Ensure the Gatch (if raised) also starts flattening when the Fowler is completely down .
Following the complete lowering of the Fowler, wait approximately 30 seconds - the time for the Fowler control
motor to reset - and verify that the motor has indeed reset itself by completely raising the Fowler using the
Fowler up control .
_____ Verify the Fowler, Gatch and Hi-Lo movements to ensure that the motion interrupt switch integrated to the
four electric actuators is operating properly .
_____ Auxiliary outlet (option available only with 120V electric system beds) working properly .
_____ Night light working properly .
_____ No cracks in the boards, siderails, wheel covers, fifth wheel hood (optional) and plastic cover of the head and
foot sections .
_____ Head end bumpers tightly secured to frame and working properly .
_____ No rips or cracks in mattress cover .
_____ Power cord not frayed .
_____ No cables worn or pinched .
_____ All electrical connections tight .
_____ All grounds secure to the frame .
_____ All casters roll properly . Check caster for cuts, wear, etc .
_____ Ground impedance not more than 100 mΩ (milliohms) .
_____ Current leakage not more than 100 µA (microamps) .
_____ Make sure that the ground chain is in place, intact and it's touching the floor (see Figure 1 .9,
_____ Check labels as specified in the Operations and Maintenance manuals for legibility, proper adherence and
integrity .