Nurse control panel (Outside siderail) (Continued)
Button name
Nurse call (optional)
K
Note: The nurse call light on the siderails show the patient which button to push to contact the nursing station in a
darkened room.
L
Level
Inside siderail light
M
Siderail bed motion
N
control lockout
O
Power LED
www.stryker.com
Operation
Activates nurse call
Raises the head end of the product or returns the product to level
Illuminates the inside siderail buttons
Five settings are available for the inside siderail lights:
Press the button once to turn on the lights at low intensity
Press the button twice for medium intensity
Press the button three times for high intensity
Press the button four times to only leave the nurse call light on
Press the button five times to turn all siderail panel lights off
Locks all siderail bed motion controls.
WARNING
Always lock the nurse control panel and patient control panel when the patient
is unattended.
To lock the patient control panel, press the lockout button once. The lockout
button illuminates when activated.
To unlock the patient control panel, press the lockout button a second time.
Illuminates when you plug the product into a hospital-grade receptacle.
Blinks when you need to replace the 9V nurse call battery.