8.2 Functional Guide
FUNCTION GUIDE
Command:
(Hand pendant) height up
Height down
Trendelenburg
Reverse trendelenburg
Lateral roll (tilt) left or right
w/rotation safety lock switch off
Rotation safety lock switch off
MIZUHOSI 2009
Result:
Both columns up
Both columns down
Head end column down and foot
end column up
Foot end column down and head
end column up
Tilt motor drives
Brake releases
28
If not, determine if
Base power is off.
Power to PC board is not
connected.
Hand pendant is non-operative,
or not plugged in.
Cable from PC board to column
is not connected, or has failed.
If none of these are the problem
then the column or the PC board
has failed.
Same as for Height Up
Same as for Height Up
Same as for Height Up
Base power is off
Power supply is not operating
correctly
Power to PC board is not
connected
Hand pendant is non-operative
Hand pendant is not plugged in
PC board is non-operative
Cabling is not connected or has
failed
Tilt motor has failed, or brake-on
sensor is closed
Brake-on sense wire is shorted
Either limit switch has failed.
Either limit switch is open
Limit switch wiring is
disconnected
Base power is off
Power to PC board is not
connected
Power supply is not operating
Cable from PC board to column
is not connected
Cable to column has failed.
Brake off switch has failed
Wire connection to switch has
failed.
Traction arc switch closed
Traction arc switch wiring
shorted
Limit switch has failed
Mechanical error
NW0498 Rev E