4.5 Tabletop Coupling Procedure
The modular table tops may be interchanged and coupled to the 5803 Advanced
Control Base as needed to provide flexibility and specific procedural capability. Position
the 5803 Base and engage all four caster locks. When using trendelenburg with tilt,
only table tops with gimbals may be used with the 5803 Base.
Figure 10: Jackson Spinal Top & Radiolucent Imaging Top with Gimbal Mounting Assemblies.
Verify that the Jackson Spinal Top or the Imaging Top being used has gimbals.
Never attach a Jackson Spinal Top or Imaging Top to the 5803 Advanced Control
Base that does not have gimbals.
Crossbar
Detail
MIZUHOSI 2009
Jackson
Spinal Top
Gimbal Coupling
Device
WARNING:
Drop Lock
Figure - T-Pin Assembly
Mounting Stud
Drop Lock
T-Pin
H-Frame
Figure 11: H-Frame Installed - Foot End Column
17
Imaging Top
Foot End
ON
H-Frame
H-Frame
Storage Bracket
T-Pin
See detail
at left
NW0498 Rev E