4.3 5803 Advanced Base Controls
The head end base of the 5803 Advanced Control Base is equipped with four indicator
lights (see figure 4). These lights are: 1) on/off AC power switch light, 2) power light, 3)
battery status light, and 4) fault light. Under normal AC operation, the on/off AC power
switch light and battery status light should be lit green and the fault light should be dark.
NOTE: If the fault light on hand pendant or base is lit, the 5803 Advanced Control Base
has detected a fault in the system. Determine fault before use (refer to sections 4.2 &
4.3).
The head end pivot assembly of the 5803 Advanced Control Base is equipped with
three indicator lights. These lights are: 1) 180° rotation lock indicator light, 2) tilt drive
status light and 3) rotation safety lock indicator light. Verify that the rotation safety lock
switch is in the on position and the 180° rotation lock handle is in the locked position.
All three lights must be illuminated before a patient is transferred to the table. These
indicator lights are shown in figure 9.
3) Rotation Safety Lock
Lighted Switch
2) Tilt Drive Status
Indicator Light
Traction Pulley
Figure 9: Head End Pivot Assembly Indicator Lights and Controls
4.3.1 Rotation Safety Lock
The rotation safety lock switch locks out the 180° rotation function. This feature
is designed to prevent unintended rotation of the table top. When the rotation
safety lock switch is illuminated, the table top can still be laterally tilted by
pressing the left or right lateral roll button on the hand pendant. The rotation
safety lock switch should be turned off only when a 180° rotation is to be
performed. If the rotation safety lock switch is not illuminated, press the rocker
switch to the on position.
MIZUHOSI 2009
1) 180° Rotation Lock
Indicator Light
13
Crossbar
180°° Rotation
Lock Lever (blue)
NW0498 Rev E